TOP-2500 Infusion Pump

You are here:


TOP Corporation’s Simple and Easy-to-Use Infusion Pump, developed for a New Standard.

More information about this product? Feel free to contact us.

TOP-2500 Infusion Pump
The TOP-2500 is a volume pump, used to administer medical fluids in a controlled and preset manner, for a broad variety of infusion therapies. It was developed as a New Standard: a very light and compact machine, with a clear focus on simplicity and safety. Preparation and use are very straightforward, following a well indicated series of successive top-to-bottom steps.

There are two modes of operation: a general mode, with use of any compatible infusion set, and the ‘H-Series’ mode, a more accurate and controlled proces with TOP’s designated infusion set.

The infusion set is guided along a slot with peristaltic fingers, through upstream and downstream occlusion sensors and an air detector sensor, safely secured behind the front door. A tube clamp prevents free flow of infusion fluid when the door is open. Bright key lights and a door light help to operate the TOP-2500 in the dark.

Safety
Events are indicated through sounds and multi-coloured lights. The internal battery ensures continued operation during power outages or transport, and may last for up until 4 hours when full.

Specifications

  • Infusion Sets: Designated Infusion Sets TOP H-Series, General 20 drops/ml Infusion Sets, General 60 drops/ml Infusion Sets (various brands)
  • Flow Rate Range Designated 20 drops/ml IV Set: 1 ~ 1.200 ml/h, Micro Mode: 0,1 ~ 99,9 ml/h
  • Flow Rate Range Designated 60 drops/ml IV Set: 1 ~ 150 ml/h, Micro Mode: 0,1 ~ 99,9 ml/h
  • Flow Rate Range General 20 drops/ml IV Set: 1 ~ 360 ml/h, Micro Mode: 1 ~ 120 drops/min.
  • Flow Rate Range General 60 drops/ml IV Set: 1 ~ 150 ml/h, Micro Mode: 1 ~ 150 drops/min.
  • Flow Rate Range Transfusion Set: 1 ~ 1.200 ml/h
  • Purge Flow Rate: Maximum flow in each flow rate range
  • KVO Flow Rate: 0,1 – 10,00 ml/h (1 ml/h steps)
  • Volume Limit Range: 1 – 9.999 ml (1 ml/h steps)
  • Total Volume Range: 0 – 9.999 ml (1 ml/h steps)
  • Infusion Accuracy Designated IV Set (H-Series): ml/h: ±5%, drops/min.: ±2%
  • Infusion Accuracy General IV Set: ml/h: ±10% (Distilled Water or Saline), drops/min.: ±2%
  • Infusion Accuracy Transfusion Set: ml/h: ±10%, drops/min.: ±2%
  • Occlusion Detection Pressure Designated IV Set (H-Series): 4 Steps Selection: PrdE 3 High (150kPa or less), PrdE 2 Medium (120kPa or less), PrdE 1 Low (80kPa or less), PrdE 0 Lowest (40kPa or less)
  • Occlusion Detection Pressure General IV Set: 2 Steps Selection: PrST 1 High (150kPa or less), PrST 0 Low (80kPa or less)
  • Air-in-Line Detection Sensitivity: 3 Steps Selection: Air 0 High, Air 1 Medium, Air 2 Low
  • Alarms: High Priority, Low Priority, Information Signals
  • Dimensions and Weight: 92 x 181 x 189 mm, approx. 1,9 kg

Special Functions Include

History, Key-lock, KVO (Keep Vein Open), Nurse Call Signal Output, Auto Power Off, Key Light, Door Light, Total Volume Display, Maintenance Timer, Battery Refresh, IV Set Limit, Standby Mode, Maintenance Mode, Buzzer Volume Setting, Buzzer Setting, Date and Time Setting, Clear Total Volume, Purge, External Communication, Anti-Bolus, Repeat Alarm

(Find the list of specifications, functions and their explanations in the TOP-2500 brochure).