TOP-5530 Syringe Pump

You are here:


Outstanding capabilities and ease of use in the clinical setting, whether on general ward or when administering an anesthetic.

More information about this product? Feel free to contact us.

TOP-5530 Syringe Pump
The TOP-5530 is a standard human-rated syringe pump with a basic set of features. The on-site clinical capabilities are outstanding, though.

Easy-to-Use, Continuous and Bolus
This ‘micro-continuous infusion pump’ — or just syringe pump — is used for the infusion of anesthetics, chemotherapeutic agents, anticancer drug, oxytocic, anticoagulants, but also for parenteral nutrition and for blood transfusions. An additional bolus can be administered during infusion, manually or with a specified volume.

Safety
Alarms are both audible and visible and they can be configured. An internal battery, which may last for up until 12 hours when full, ensures continued infusion during power outages or transport. An incorporated Drugs Library (available with External Communication Connector only) is there to help find the appropriate infusion settings.

Specifications

  • Compatible Syringes: 10, 20, 30, 50 ml (various brands)
  • Flow Rate Range: 0,10 ~ 1.200 ml/h (0,1 ml/h increments) depending on syringe size
  • Agent Range: 0,1 ~ 1.000 mg
  • Solution Range: 0,1 ~ 1.000 ml
  • Weight Range: 0,1 ~ 300 kg
  • KVO Rate Range: 0,01 – 99,99 µg/kg/min, 0,01 – 99,99 mg/kg/h
  • Bolus Rate: 300 to 1.200 ml/h depending on syringe size
  • Bolus Volume: 0,1 – 50,0 ml (0,1 ml steps), 0,1 – 50.000,0 μg (0,1 μg steps), 0,1 – 5.000,0 mg (0,1 mg steps)
  • KVO Rate Range: 0,1 ~ 5,0 ml/h (0.1 ml/h step)
  • Volume Limit Setting: 0,1 ~ 1.000,0 ml
  • Total Volume Display Range: 0,0 ~ 1.000,0 ml
  • Infusion Accuracy: Mechanical ±1%, including syringe ±3%
  • Occlusion Detection: High, Middle, Low, Very Low
  • Alarms: High Priority, Low Priority, Information Signals
  • Dimensions and Weight: 320 x 90 x 160 mm, 2,0 kg

Functions Include

Auto Power Off, Alarm Reminder, Power Supply Switch Alarm, KVO (Keep Vein Open), Standby, Buzzer Setting, History, Key Lock, Syringe Brand Limit, Volume Limit Display Switching, battery Refresh, Maintenance Timer, Flow Rate Upper Limit Value Setting, User Syringe Registration, Drug Library

(Find a full list of specifications, functions and their explanations in the TOP-5530 brochure).